Protocol Settings
R&S
®
EX-IQ-BOX
125
User Manual 1409.5570.32 ─ 05
C&M Data
Defines the number of the Control & Management bit word as defined in the C&M Size
parameter. This setting can be specified in transmitter mode, in receiver mode it is read-
only.
Note:
For this setting, the SCPI notation lightly differs between the command for the R&S
signal generators, R&S
DigIConf and the command for the R&S analyzers.
Remote command:
Late Frame Sync
Defines if the sync signal appears at the first period of a data bit (parameter set to On)
or if the sync signals appears one period before (parameter is Off).
Remote command:
Gain File
Defines a binary file contianing a gain table. For the format of the gain table, refer to the
appropriate generator section.
Note:
A gain file can be selected in input / receiver mode and for "C&M" >=8 !
Remote command:
Enable Automatic Gain
Activates the automatic gain handling.
Note:
A gain file can be used in input / receiver mode and for "C&M" >=8 !
Remote command:
7.1.11 SPI Setup
The "SPI Setup" panel contains clock parameters for data transmission of SSI (
S
erial
S
ynchronuos
I
nterface) signals.
Clock Frequency
Sets a system clock frequency. Available freqencyies are 1
MHz, 2.5
MHz, 5
MHz,
8.333
MHz, 12.5
MHz or 25
MHz.
Remote command:
Clock Phase
Selects 0 or 180 degree clock phase.
Remote command:
User Defined