Remote Control Commands
R&S
®
EX-IQ-BOX
293
User Manual 1409.5570.32 ─ 05
[:SOURce<hw>]:EBOX:CPRI:CM:ETH:PTR?
The command queries the resulting Ethernet pointer after the link setup, i.e. the position
after the handshake between the R&S EXBOX-B85 and the DUT.
Return values:
<Pptr>
float
Range:
20 to 63
Increment: 1
*RST:
20
Example:
SOUR1:EBOX:CPRI:CM:ETH:PTR?
check the position of the ethernet pointer after link setup.
Response:
45
the resulting pointer is at position 45.
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:EBOX:CPRI:CM:ETH:STATe
<State>
The command activates the C&M data exchange via the Ethernet interface.
Parameters:
<State>
OFF | ON
*RST:
OFF
Example:
SOUR1:EBOX:CPRI:CM:ETH:STAT ON
switches on fast C&M.
Manual operation:
See
[:SOURce<hw>]:EBOX:CPRI:CM:HDLC:ACTive?
The command queries whether the interface is active.
Return values:
<Active>
OFF | ON
*RST:
OFF
Example:
SOUR1:EBOX:CPRI:CM:HDLC:ACT?
checks the activity of the HDLC interface.
Response:
0
the interface is not active.
Usage:
Query only
Manual operation:
See
CPRI Remote Control Commands