Remote Control Commands
R&S
®
EX-IQ-BOX
281
User Manual 1409.5570.32 ─ 05
[:SOURce<hw>]:EBOX:USER:TEST:TX:SINE:FREQuency
<Frequency>
Sets the frequency of the sine test signal. The frequency is limited to "0.4 f
s
", and the
sampling frequency "f
s
" depends on the clock, data rate and the interleaving mode set-
tings.
Parameters:
<Frequency>
float
Range:
-4 MHz to 4 MHz
Increment: 1 kHZ
Example:
SOUR1:EBOX:USER:DIR TRAN
SOUR1:EBOX:USER:TEST:TX:SINE:FREQ 100KHZ
sets to "Transmitter" and the signal frequency "100 kHz".
Manual operation:
See
[:SOURce<hw>]:EBOX:USER:TEST:TX:STATe
<State>
Switches the test generator on or off. While the test generator is switched on, the test
signal replaces the I/Q data source coming from "DIG IQ IN".
Parameters:
<State>
OFF | ON
*RST:
OFF
Example:
SOUR1:EBOX:USER:DIR TRAN
SOUR1:EBOX:USER:TEST:TX:STAT ON
sets "Transmitter" and switches on the test generator.
Manual operation:
See
9.3.8 Trigger Subsystem
The
USER:TRIGger
subsystem contains the commands for configuring a trigger signal
when the R&S
EX-IQ-BOX is working in transmitter mode.
....................................................................281
................................................................282
.................................................................282
..............................................................282
..............................................................282
...............................................................283
[:SOURce<hw>]:EBOX:USER:TRIGger:ARM
Stops the current signal output. It continues either automatically or synchronized, accord-
ing to the used trigger source.
Example:
EBOX:USER:TRIG:ARM
stops signal output.
User Defined - Remote Control Commands