![Rohde & Schwarz EX-IQ-BOX User Manual Download Page 116](http://html1.mh-extra.com/html/rohde-and-schwarz/ex-iq-box/ex-iq-box_user-manual_1477982116.webp)
Protocol Settings
R&S
®
EX-IQ-BOX
101
User Manual 1409.5570.32 ─ 05
Fig. 7-4: User defined > Protocol dialog
Format
Sets the transmission protocol of the R&S
EX-IQ-BOX for receiving or sending a base-
band signal from or to an external device (DUT) via the user interface of the R&S
EX-IQ-
BOX.
"Serial"
Tranmsits each sample serially, i.e. bitwise from the DUT to the
R&S
EX-IQ-BOX or vice versa.
Example:
The following graph shows an incoming serial I/Q signal from the DUT:
Fig. 7-5: User defined > I/Q IN serial signal flow
Direction = Receiver
Format
= Serial
In serial mode, I data and Q data are sent via the input and output pins
IO and QO, respectively.
User Defined