Risk of hardware damage
Digital, electrical signals base on various logical technologies, such as TTL or CMOS with
several voltage levels. Therefore, the signal settings in the R&S
EX-IQ-BOX must cor-
respond to the signal, in both, the logical standard as well as the voltage value.
Any deviations may damage the DUT or the R&S
EX-IQ-BOX.
A change of the logic type during the operation is performed only when you confirm the
following warning message:
●
Check the voltage values of your test setup.
●
Select "OK" if the correct logic type is selected.
●
Select "Cancel" if an incompatible logic type is selected. Adjust your setup accord-
ingly and set correct logic type in the main configuration dialog for User Defined pro-
tocols, see
6.2.1.1
Connection
► Establish the USB and LVDS connection between the R&S
EX-IQ-BOX and the R&S
signal generator, as described in
chapter 4.5, "Connecting the R&S EX-IQ-BOX"
on page 42.
The R&S
EX-IQ-BOX can be connected or disconnected while the R&S instrument
is in operating mode.
6.2.1.2
Indication of the R&S EX-IQ-BOX in an R&S Signal Generator
The graphical user interfaces (GUIs) of the signal generators R&S AMU200A, R&S
SMU200A and R&S SMJ100A display the architecture and the signal flow in a block
diagram. If an R&S
EX-IQ-BOX is connected, the R&S instrument recognizes this device
automatically, and marks the connection in the block diagram accordingly.
Configuration via R&S Instruments