![Rohde & Schwarz CMW500 Series User Manual Download Page 223](http://html1.mh-extra.com/html/rohde-and-schwarz/cmw500-series/cmw500-series_user-manual_1477718223.webp)
Command Reference
R&S
®
CMW500
223
User Manual 1173.9463.02 ─ 02
For a description of the variables <netw_std>, <func_grp> and <appl> refer to
.
Return values:
<ConditionBits>
Range:
0 to 65535 (decimal representation)
Example:
STAT:OPER:TASK:A:GPRF:GEN:UNIV:COND?
Query the CONDition part of the GPRF generator status register
to check the current generator state.
Usage:
Query only
Firmware/Software:
V1.0.4.11
STATus:OPERation:ENABle
<MaskBits>
STATus:OPERation:TASK:ENABle
<MaskBits>
STATus:OPERation:TASK:A:ENABle
<MaskBits>
STATus:OPERation:TASK:A:<netw_std>:ENABle
<MaskBits>
STATus:OPERation:TASK:A:<netw_std>:<func_grp>:ENABle
<MaskBits>
STATus:OPERation:TASK:A:<netw_std>:<func_grp>:<appl>:ENABle
<MaskBits>
Sets the enable mask which allows true conditions in the EVENt part of the status register
to be reported in the summary bit. If a bit is 1 in the enable register and its associated
event bit transitions to true, a positive transition will occur in the summary bit reported to
the next higher level. See also
Structure of an SCPI Status Register
For a description of the variables <netw_std>, <func_grp> and <appl> refer to
.
Parameters:
<MaskBits>
Range:
0 to 65535 (decimal representation)
Example:
STAT:OPER:TASK:A:GPRF:MEAS:ENAB 1536
Set bits no. 9 and 10 of the ENABle part of the status register for
GPRF measurements (1536 = 512 + 1024 = 2
9
+ 2
10
).
Firmware/Software:
V1.0.4.11
STATus:OPERation:ESRQ
<MaskBits>
STATus:OPERation:TASK:ESRQ
<MaskBits>
STATus:OPERation:TASK:A:ESRQ
<MaskBits>
STATus:OPERation:TASK:A:<netw_std>:ESRQ
<MaskBits>
STATus:OPERation:TASK:A:<netw_std>:<func_grp>:ESRQ
<MaskBits>
STATus:OPERation:TASK:A:<netw_std>:<func_grp>:<appl>:ESRQ
<MaskBits>
Sets the enable mask for the register to the specified value. Also sets the relevant bit in
the enable mask of all higher registers up to the STATus:OPERation register and the
SRE. Thus the entire reporting path from the register up to the SRE is enabled so that
an SRQ can be generated.
If the enable mask is set to 0, the higher registers are not modified.
For a description of the variables <netw_std>, <func_grp> and <appl> refer to
.
Parameters:
<MaskBits>
Range:
0 to 65535 (decimal representation)
Instrument-Control Commands