![Rohde & Schwarz CMW500 Series User Manual Download Page 201](http://html1.mh-extra.com/html/rohde-and-schwarz/cmw500-series/cmw500-series_user-manual_1477718201.webp)
Command Reference
R&S
®
CMW500
201
User Manual 1173.9463.02 ─ 02
Parameters:
'<BufferLabel>'
String parameter, used to identify the buffer in all buffer commands
<LineNo>
Line of results in the buffer
Example:
See
Usage:
Query only
Firmware/Software:
V2.1.10
FETCh:BASE:BUFFer:LINecount?
'<BufferLabel>'
Returns the size (number of lines) of a buffer.
Parameters:
'<BufferLabel>'
String parameter, used to identify the buffer in all buffer commands
Example:
See
Usage:
Query only
Firmware/Software:
V2.1.10
STARt:BASE:BUFFer
'<BufferLabel>'
Creates and activates a buffer. If the buffer exists already, it is cleared (equivalent to
).
Parameters:
'<BufferLabel>'
String parameter, used to identify the buffer in all buffer commands
Example:
See
Usage:
Event
Firmware/Software:
V2.1.10
STOP:BASE:BUFFer
De-activates the active buffer. Only one buffer can be active at one time. The buffer and
buffer contents are maintained, however, data recording is discontinued. Use
to re-activate a buffer.
Example:
See
Usage:
Event
Firmware/Software:
V2.1.10
7.3.3 Frequency-Dependent Correction Settings
The following commands administrate and activate/deactivate correction tables for fre-
quency-dependent attenuation/gain. For additional information concerning the usage of
Instrument-Control Commands