![Rohde & Schwarz CMW500 Series User Manual Download Page 265](http://html1.mh-extra.com/html/rohde-and-schwarz/cmw500-series/cmw500-series_user-manual_1477718265.webp)
Command Reference
R&S
®
CMW500
265
User Manual 1173.9463.02 ─ 02
Parameters:
<Direction>
IN | OUT
IN:
Input connector
OUT:
Output connector
*RST:
IN
Example:
TRIGger:BASE:EXTB:DIRection OUT
Use TRIG B as an output connector.
Firmware/Software:
V1.0.4.11
TRIGger:BASE:EXTA:SOURce
'<Source>'
TRIGger:BASE:EXTB:SOURce
'<Source>'
Selects the output trigger signals to be routed to the TRIG A and TRIG B connectors. The
signals listed below are always available. Depending on the installed options additional
values may be available. A complete list of all supported values can be retrieved using
TRIGger:...:CATalog:SOURce?.
Parameters:
'<Source>'
'No Connection':
No output trigger signal available
'Base1: External TRIG A':
Input trigger signal from the TRIG A
connector
'Base1: External TRIG B':
Input trigger signal from the TRIG B
connector
*RST:
'No Connection'
Example:
TRIGger:BASE:EXTB:SOURce "Base1: External TRIG
A"
Select the input signal from TRIG A as an output trigger signal for
connector TRIG B.
Firmware/Software:
V1.0.4.11
Instrument-Control Commands