Basic Instrument Functions
R&S
®
CMW500
116
User Manual 1173.9463.02 ─ 02
Fig. 5-22: Blockview dialog - Selectable Routing
Note that each drop-down menu to the left lists only the installed firmware applications
supported by the block. The first (uppermost) "BB Generator" supports the first instance
of generator applications, the second "BB Generator" the second instance. The same
applies to "BB Measurement" for measurement applications. Signaling applications
require a certain type of signaling unit - either universal or wideband.
The following actions are possible in "Show Selectable Routing" mode.
Displaying the signal path configured for an application
► Use the drop-down menus to the left to select the firmware application.
The signal path currently configured for the selected application is displayed and
highlighted. This action has no impact on the allocation of resources, especially of
baseband boards or signaling units. It only shows the signal path that would be used
if the application would be running.
Highlighting the signal path of one block / application
► Select a block to the left.
The currently selected block is marked by a border and the related signal path is
highlighted. This is especially useful if you have selected applications in several drop-
down menus, so that the segments of several paths are shown in parallel.
Blockview Dialog