Remote Control
R&S
®
CMW500
156
User Manual 1173.9463.02 ─ 02
All instances can access both frontends. The following table lists the supported parameter
values and used *RST values for selection of RF connectors and TX/RX boards.
Instance 1
Instance 2
Input connector
RF1C | RF2C | RF3C | RF4C | RFAC | RFBC
Output connector
RF1C | RF1O | RF2C | RF3C | RF3O | RF4C | RFAC |
RFAO | RFBC
Mapping
RFAC, RFAO, RFBC
RF1C, RF1O, RF2C
TX module
TX1
(for RF1/RF2 connectors only)
TX2
(for RF3/RF4 connectors only)
RX module
RX1
(for RF1/RF2 connectors only)
RX2
(for RF3/RF4 connectors only)
*RST connectors
RF1C
RF3C
*RST TX/RX
TX1, RX1
TX2, RX2
6.4.4.5
Two RF Paths, Two Frontends, Fixed Link, One Sub-Instrument
This section applies to an instrument with two RF paths, two frontends, a fixed link (R&S
CMW-S550A) and two instances per firmware application. All resources are grouped in
one sub-instrument.
Each instance is fix assigned to one frontend. The following table lists the supported
parameter values and used *RST values for selection of RF connectors and TX/RX
boards.
Instance 1
Instance 2
Input connector
RF1C | RF2C | RFAC |
RFBC
RF3C | RF4C | RFAC |
RFBC
Output connector
RF1C | RF1O | RF2C | RFAC
| RFAO | RFBC
RF3C | RF3O | RF4C | RFAC
| RFAO | RFBC
Mapping
RFAC, RFAO, RFBC
RF1C, RF1O, RF2C
RF1C, RF1O, RF2C
Virtual connector names cannot
be used (result in a settings con-
flict).
TX module
TX1
TX2
RX module
RX1
RX2
*RST connectors
RF1C
RF3C
*RST TX/RX
TX1, RX1
TX2, RX2
6.4.4.6
Two RF Paths, Two Frontends, Two Sub-Instruments
This section applies to an instrument with two RF paths, two frontends and two instances
per firmware application. The instrument is split into two sub-instruments.
Control of the Instrument