![Rohde & Schwarz CMW500 Series User Manual Download Page 50](http://html1.mh-extra.com/html/rohde-and-schwarz/cmw500-series/cmw500-series_user-manual_1477718050.webp)
Getting Started
R&S
®
CMW500
50
User Manual 1173.9463.02 ─ 02
Fig. 3-1: General Purpose RF Generator dialog
2. Click "Routing > Connector" and select your RF output connector. In the following we
assume that RF1 COM is used.
3. Select the "Frequency" (1200 MHz) and "Level (RMS)" (–30 dBm) of the RF output
signal.
4. If your test setup contains a known, frequency-independent attenuation, enter the
value as an "External Attenuation (Output)".
5. Ensure that the "List Mode" is disabled (Off).
6. Select "Baseband Mode: Dual Tone".
7. In the "Baseband Configuration > Dual Tone" section, configure the properties of the
dual-tone signal. To superimpose two CW signals at different frequencies and levels,
set the "Offset Frequency" of both signals and define a "Ratio (Source 1 - Source 2)
that is different from 0 dB; see figures above and below.
8. Press ON | OFF to switch the RF generator on.
Sample Session