Command Reference
R&S
®
CMW500
263
User Manual 1173.9463.02 ─ 02
Suffix:
<inst>
.
1..2
Selects the sub-instrument
Parameters:
<StartMode>
AUTO | EXPLicit
AUTO
: Start tracing automatically when the instrument is started
EXPLicit
: Start tracing via the command
Default value: EXPLicit
Example:
TRACe:REMote:MODE:FILE1:STARtmode AUTO
Enable automatic start of tracing for the next (re-)start of the
instrument.
Firmware/Software:
V2.0.10
TRACe:REMote:MODE:FILE<inst>:STOPmode
<StopMode>
Specifies how / when tracing shall be stopped and the trace file be closed.
Suffix:
<inst>
.
1..2
Selects the sub-instrument
Parameters:
<StopMode>
AUTO | EXPLicit | ERRor | BUFFerfull
AUTO
: Stop tracing automatically when the instrument is shut
down
EXPLicit
: Stop tracing via the command
ERRor
: Stop tracing when an SCPI error occurs
BUFFerfull
: Stop tracing when the maximum file size is reached
Default value: EXPLicit
Example:
TRACe:REMote:MODE:FILE1:STOPmode BUFFerfull
Stop tracing when the maximum file size is reached (sub-instru-
ment 1).
Firmware/Software:
V2.0.10
TRACe:REMote:MODE:DISPlay:ENABle
<Enable>
Enables or disables the display of the SCPI remote trace. Two modes are available when
the display is enabled: a live mode and an analysis mode.
Parameters:
<Enable>
ANALysis | LIVE | OFF
ANALysis
: stop tracing to analyse already traced messages
LIVE
: trace messages and display them
OFF
: disable the report display
Default value: OFF