![Rohde & Schwarz CMW500 Series User Manual Download Page 175](http://html1.mh-extra.com/html/rohde-and-schwarz/cmw500-series/cmw500-series_user-manual_1477718175.webp)
Remote Control
R&S
®
CMW500
175
User Manual 1173.9463.02 ─ 02
●
STATus:OPERation:TASK:A:GPRF:GENerator2:UNIVersal:EVENt?
etc.
Waiting until a certain state is reached or a certain state transition occurrs is possible
using the following commands:
●
STATus:OPERation:TASK:A:GPRF:GENerator1:UNIVersal:WCONdition?
etc.
●
STATus:OPERation:TASK:A:GPRF:GENerator2:UNIVersal:WEVent?
etc.
These commands are also available as extended versions, providing more comfort:
●
...:XENable
,
...:XESRq
,
...:XNTRansition
,
...:XPTRansition
●
...:XCONdition
,
...:XEVent
●
...:XWCondition
,
...:XWEVent
More sophisticated evaluations of the hierarchy are possible using e.g.:
●
STATus:CONDition:BITS:ALL?
●
STATus:EVENt:BITS:ALL?
●
STATus:MEASurement:CONDition:RDY?
etc.
For command descriptions refer to:
●
chapter 7.3.7.2, "STATus:OPERation (Elementary Commands)"
●
chapter 7.3.7.3, "STATus:OPERation (Extended Commands)"
●
chapter 7.3.7.4, "STATus:OPERation (Overall Evaluation)"
6.6.3.5
STATus:QUEStionable
The STATus:QUEStionable register indicates whether the data currently being acquired
is of questionable quality.
The R&S CMW500 does not use the STATus:QUEStionable register.
6.6.4 Application of the Status Reporting System
The purpose of the status reporting system is to monitor the status of one or several
devices in a measuring system. To do this and react appropriately, the controller must
receive and evaluate the information of all devices. The following standard methods
described in the following sections are used:
●
Service request (SRQ) initiated by the measuring device
●
Serial poll of all devices in the bus system, initiated by the controller in order to find
out who sent a SRQ and why
●
Parallel poll of all devices
●
Query of a specific instrument status by means of commands
●
Query of the error queue
Status Reporting System