Remote Control
R&S
®
CMW500
144
User Manual 1173.9463.02 ─ 02
Simplified statistics
Some measurement contexts provide simplified statistical settings. Refer to the mea-
surement and remote control description for details.
6.4.2.4
Retrieving Measurement Results
The results of a measurement can be retrieved by means of
FETCh...?
or
READ...?
queries. Limit check results are retrieved by
CALCulate...?
queries. All command
types have the same structure:
FETCh:<FWA>:<Context>:<View>:<Statistics>...?
READ:<FWA>:<Context>:<View>:<Statistics>...?
CALCulate:<FWA>:<Context>:<View>:<Statistics>...?
e.g.
FETCh:GPRF:MEASurement<i>:POWer:CURRent?
The three identifiers
<FWA>, <Context>, <View>
and
<Statistics>
denote the
Measurement Contexts and Views
, and the
respectively.
<View>
is only required for measurement contexts providing several views.
The number and type of measurement results depends on the measurement context and
view; refer to the relevant reference documentation.
FETCh...? Command
Waits until the end of the current measurement cycle (if the measurement is running) and
returns the results including the
FETCh...?
is similar to the
, however, it does not start a new measurement. The measurement
must have been started (
INITiate...
; see
FETCh...?
command can be executed.
Measurement states
A
FETCh...?
command can be used in the RUN state as described above or in the RDY
state, provided that the R&S CMW500 has stored valid results. If no valid results are
available, NCAP, NAV or INV are returned for each unavailable valid result, see also
Return Values NCAP, NAV and INV
The following table gives an overview of the behavior depending on the measurement
state.
State
Valid results?
Effect of FETCh...?
OFF
No
FETCh...?
should not be used in the OFF state as valid results are
never available.
RUN
Yes/No
R&S CMW500 waits until the current measurement cycle is complete
and returns the results. In continuous repetition mode the R&S
CMW500 continues acquiring results without reaching the RDY state.
RDY
Yes/No
Returns the results immediately.
Control of the Instrument