Command Reference
R&S
®
CMW500
243
User Manual 1173.9463.02 ─ 02
Example:
SYSTem:BASE:DEVice:SUBinst?
Returns
0,1
or
0,2
or
1,2
.
Usage:
Query only
Firmware/Software:
V2.0.10
SYSTem:DEVice:ID?
Queries the device identification (independent of sub-instruments).
Return values:
'<DeviceID>'
Device ID string.
Usage:
Query only
Firmware/Software:
V1.0.10.1
7.3.8.2
Display Settings
The following commands configure the appearance of the Graphical User Interface (GUI).
...................................................................................243
..............................................................................................243
SYSTem:BASE:DISPlay:FONTset
<FontSize>
Selects the font size for the GUI labels.
Parameters:
<FontSize>
DEF | LRG
DEF:
Small fonts
LRG:
Large fonts
*RST:
DEF
Example:
SYSTem:BASE:DISPlay:FONTset LRG
Select large fonts.
Firmware/Software:
V1.0.5.3
SYSTem:DISPlay:UPDate
<Enable>
Defines whether the display shall be updated or not while the instrument is in the remote
state. If the display (update) is switched off, the normal GUI is replaced by a static image
while the instrument is in the remote state. Switching off the display can speed up the
measurement. This is the recommended state.
See also:
Instrument-Control Commands