![Rohde & Schwarz CMW500 Series User Manual Download Page 228](http://html1.mh-extra.com/html/rohde-and-schwarz/cmw500-series/cmw500-series_user-manual_1477718228.webp)
Command Reference
R&S
®
CMW500
228
User Manual 1173.9463.02 ─ 02
Example:
STAT:OPER:BIT9:ENAB 1
Set bit no. 9 of the ENABle part of the STATus:OPERation regis-
ter.
Firmware/Software:
V1.0.4.11
7.3.7.3
STATus:OPERation (Extended Commands)
The commands listed in this section serve the same purpose as the elementary com-
mands, see
chapter 7.3.7.2, "STATus:OPERation (Elementary Commands)"
on page 221. However the extended commands provide more comfort by mapping bit
values to the corresponding status register mnemonics whenever possible. Thus they
allow to control the STATus:OPERation register hierarchy without knowing it by heart at
bit level.
For a complete overview of the mnemonics used in the register hierarchy see
and
Settings
Assume that you want to set the enable mask for the states OFF and RDY of the GPRF
external power sensor measurement 1, up to the highest level of the hierarchy. You can
do this by studying the hierarchy and setting the correct bits using the following elemen-
tary commands:
STATus:OPERation:TASK:A:GPRF:MEAS1:EPSensor:ENABle 9
STATus:OPERation:TASK:A:GPRF:MEAS1:ENABle 4
STATus:OPERation:TASK:A:GPRF:ENABle 1
STATus:OPERation:TASK:A:ENABle 2
STATus:OPERation:TASK:ENABle 1
STATus:OPERation:ENABle 512
Or you use the following extended commands. Knowing the first command is sufficient
to write down all subsequent commands correctly, because the first (lowest level) com-
mand reflects the entire register path up to the highest level and contains all mnemonics:
STATus:OPERation:TASK:A:GPRF:MEAS1:EPSensor:XENable (OFF,RDY)
STATus:OPERation:TASK:A:GPRF:MEAS1:XENable (EPSensor)
STATus:OPERation:TASK:A:GPRF:XENable (MEAS1)
STATus:OPERation:TASK:A:XENable (GPRF)
STATus:OPERation:TASK:XENable (A)
STATus:OPERation:XENable (TASK)
Extended commands accept also decimal numbers in addition to mnemonics. If you use
a mixture of numbers and mnemonics, both the decimal numbers and the mnemonics
are internally translated into bits and the sum of all bits is set. Example: To set the enable
mask for the GPRF measurements POWer (bit 0), EPSensor (bit 2) and IQRecorder (bit
3) you can e.g. use one the following commands. All commands yield the same result.
STATus:OPERation:TASK:A:GPRF:MEAS1:XENable (POWer,EPSensor,IQRecorder)
STATus:OPERation:TASK:A:GPRF:MEAS1:XENable (POWer,EPSensor,8)
STATus:OPERation:TASK:A:GPRF:MEAS1:XENable (POWer,EPSensor,12)
Instrument-Control Commands