SYSTem:HELP:STATus[:REGister]?
Returns a list of paths for the STATus:OPERation registers.
Each path is represented by a string containing all registers from highest level down to
the individual register, separated by colons.
For the GPRF power measurement for example the following paths are listed:
"STATus:OPERation", "STATus:OPERation:TASK", "STATus:OPERation:
TASK:A", "STATus:OPERation:TASK:A:GPRF", "STATus:OPERation:TASK:
A:GPRF:MEASurement", "STATus:OPERation:TASK:A:GPRF:MEASurement:
POWer"
Usage:
Query only
Firmware/Software:
V2.0.10
SYSTem:KLOCk
<State>
Locks or unlocks the local controls of the instrument. This includes the front panel keys,
the keyboard, or other local interfaces.
Parameters:
<State>
ON | OFF
ON:
Local key locked (key lock enabled)
OFF:
Local keys unlocked
*RST:
OFF
Example:
SYSTem:KLOCk ON
Lock the local keys.
Usage:
SCPI confirmed
Firmware/Software:
V1.0.0.4
SYSTem:PRESet
SYSTem:PRESet:BASE
SYSTem:PRESet:ALL
SYSTem:RESet
SYSTem:RESet:BASE
SYSTem:RESet:ALL
A PRESet sets the parameters to values suitable for local/manual interaction, a RESet
to values suitable for remote operation.
The commands reset/preset the entire R&S CMW500 or only certain parts:
●
SYSTem:PRESet/RESet:
current sub-instrument
●
...:BASE:
base settings, e.g. external trigger settings in setup dialog
●
...:ALL:
all sub-instruments and base settings
The command
*RST
is synonymous to
SYSTem:RESet
.
Instrument-Control Commands