![Rohde & Schwarz CMW500 Series User Manual Download Page 167](http://html1.mh-extra.com/html/rohde-and-schwarz/cmw500-series/cmw500-series_user-manual_1477718167.webp)
Remote Control
R&S
®
CMW500
167
User Manual 1173.9463.02 ─ 02
All status registers have the same internal structure, see
For more information on the individual status registers see
SRE register
The service request enable register SRE can be used as ENABle part of the STB if the
STB is structured according to SCPI. By analogy, the ESE can be used as the ENABle
part of the ESR.
6.6.1 Overview of Status Registers
The status registers of the R&S CMW500 are implemented as shown below.
6.6.2 Structure of an SCPI Status Register
Each standard SCPI register consists of 5 parts which each have a width of 16 bits and
have different functions. The individual bits are independent of each other, i.e. each
hardware status is assigned a bit number which is valid for all five parts. Bit 15 (the most
Status Reporting System