Roche Diagnostics
B-156
Operator’s Manual · Version 3.1
7 Configuration
Cedex Bio System
Applications
Laboratory correlations (ratio applications)
Reference range definitions (ratio applications)
Result conversion definitions
Item
Values
Comment
Factor
Used for sample and control results.
Final result = (result *
Factor
) +
Offset
The correlation factor is used to compensate for
differences in measuring methods of different
instruments. By applying a factor, the final
results of the different instruments can be
directly compared.
Offset
The offset is used to compensate for differences
in measuring methods of different instruments.
By adding the offset, the final results of the dif-
ferent instruments can be directly compared.
Used
On
,
Off
Choose
On
to ensure that the defined factor and
offset are applied the final ratio result.
Table B-17
Laboratory correlation definitions for ratio applications
Item
Values
Comment
Valuation
No Valuation
No valuation is performed by the system.
Reference Range
Use the limits defined as
Low Limit
and
High
Limit
.
Low Used
On
,
Off
Refers to
Low Limit
.
High Used
On
,
Off
Refers to
High Limit
.
Low Limit
Values in standard
units.
Results that are outside the limit are flagged.
High Limit
Values in standard
units.
Results that are outside the limit are flagged.
Table B-18
Reference range definitions for ratio applications
Item
Values
Comment
Standard Unit
Laboratory Unit
Alphanumeric
characters
Type the unit.
Factor
Conversion factor for converting results from
one unit to another, for example from mg/dL to
mmol/L.
Displayed Unit
Standard Unit
Laboratory Unit
Dec. Position
0 ... 4
Number of digits after decimal point that are
displayed on the screens.
Table B-19
Result conversion definitions
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...