10 Result flags
List of flags
? Cal
Meaning
A result where its calibration is flagged.
Message ID
110
Priority
29
Possible cause
The calibration was in question and has been flagged, however the calibration could
be used to calculate the result (compare with Cal Error where the calibration did not
provide a usable result).
Recommended actions
Check and redo calibration if needed.
Cal Error
Meaning
No valid calibration data available.
Message ID
43
Priority
24
Possible cause
Caused by an alarm that occurred during calibration because the calibration could
not provide a usable result.
Recommended actions
Check the flags of the calibration and proceed to deal with the calibration first.
Calc Error
Meaning
Calculation error.
o
General calculation error.
o
Slope or nonlinear standard curve cannot be calculated due to a calibration error.
Message ID
16
Priority
13
Possible cause
o
Calibrator outdated or deteriorated.
o
Misplacement of the calibrator tubes.
Recommended actions
o
Repeat calibration with fresh calibrators.
o
If the calibrators were incorrectly positioned, replace the calibrators and repeat
the calibration.
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...