6 Special operations
Cedex Bio System
Preparing a new disk
Loading reagent sets
9
Press .
A screen is displayed, asking you to scan the bottle barcode.
10
Scan the barcode on the bottle.
11
Remove the cap from the bottle and place it in the reagent bay.
12
Press
to confirm the insertion.
If you confirm without placing the bottle, the system assumes that the bottle is
placed.
If you press
after placing the bottle, the position is deemed empty.
When the first reagent bottle of a set is loaded, the reagent set is defined. From
this moment on, the reagents are handled as part of the set. You no longer handle
them as individual reagents.
13
Do one of the following:
14
Press .
15
On the disk overview screen, press
to load the next reagent. Proceed as
described in steps 10 through 15.
If
Do this
There is another bottle
belonging to the set:
A screen is displayed, asking you to scan the bottle.
1. Scan the bottle
2. Insert the bottle.
3. Press
to confirm the insertion.
All bottles of the set are
inserted:
1. Close the main cover.
On the screen with the reagent set buttons, the button
for the new set is now present.
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...