Transfer head arrest
When the transfer head is obstructed in its horizontal movement, it immediately
stops. All pipetting and processing actions stop.
Mixing motor
The mixing motor is mounted on the carriage. When running, it generates a circular
movement of the probe. This movement is used for mixing the content of cuvettes
and reagent bottles.
Probe
The probe has a flat tip. This is required for tube bottom detection. Because such a
probe cannot pierce a bottle cap, all bottles must be placed on the instrument with
their caps removed.
Level detection
A sensor detects when the probe enters a fluid. On the basis of this level, the system
establishes whether there is enough fluid to perform the scheduled pipetting action.
Tube Bottom detection
A physical sensor is activated as soon as the probe touches the bottom of a sample
tube.
This mechanism also works when the probe touches an object outside the tube. In
both cases, probe action stops and an appropriate alarm message is generated.
A
Mixing motor
B
Transfer Z motor
C
Probe holder
D
Probe
E
Mixer movement axis
Figure A-16
Transfer head
A
B
D
E
C
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...