Roche Diagnostics
B-38
Operator’s Manual · Version 3.1
5 Daily operation
Cedex Bio System
Analyzing samples
Defining orders
Preconditions
o
All preparation tasks are complete.
o
The required tests are installed on the system and ready for use (calibration and
QC performed).
o
The system status is either
Standby
or
Operating
.
o
At least one sample position on the sample area is free.
o
There can be only one order per test and sample.
o
Possible post-dilution is predefined in the application definition.
WARNING
Incorrect results due to declining sample quality
Evaporation of sample fluid may lead to incorrect results. In ambient temperatures of more
than 25°C, be sure to start processing immediately after placing the sample and defining
the order. When processing of the order is finished, be sure to immediately remove the
sample from the sample area.
WARNING
Incorrect results due to insufficient fluid
Insufficient fluid may lead to inaccurate pipetting and consequently to incorrect results.
Always fill the tubes with enough fluid that at least the defined dead volume of fluid is left
when pipetting is complete.
e
Incorrect results due to inappropriate tube and cup placement
Inappropriate tube and cup placement may lead to inaccurate pipetting and consequently
to incorrect results.
Make sure that the primary tubes are placed centrally and perfectly vertically in the holders
in the sample area and that they are inserted firmly.
Make sure that secondary tubes are placed centrally on the primary tubes and that they
rest fully on them.
NOTICE
Probe damage and instrument malfunction due to not removing primary tube caps
The probe is not designed to pierce tube caps. It can get damaged when trying to pierce
tube caps, which can lead to instrument malfunction.
Always remove the caps of primary tubes before placing them on the instrument.
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...