Roche Diagnostics
Operator’s Manual · Version 3.1
B-27
Cedex Bio System
5 Daily operation
Preparing the system
13
Remove the cap from the bottle and place it in the reagent bay.
14
Press
to confirm the insertion.
If you confirm without placing the bottle, the system assumes that the bottle is
placed.
If you press
after placing the bottle, the position is deemed empty.
When the first reagent bottle is loaded, the reagent set is defined. From this
moment on, the reagents are handled as part of the set. You no longer handle
them as individual reagents.
15
Do one of the following:
16
On the screen with the reagent set buttons, press the button for the set you just
inserted.
A screen is displayed that shows details of the set.
The status description points to the action you need to take. If you just inserted a
reagent set, you may have to perform the initial calibration and QC.
Time limit for placing reagent bottles
o
The system assumes that you place the bottle that you just scanned.
o
You need to place the reagent bottle on the reagent disk within 15 seconds of
scanning the reagent barcode. Failing to confirm the placement within 15 seconds
cancels the current identification process. You will be asked to scan the bottle
again.
WARNING
Incorrect results due to not placing the identified reagent
The system assumes that the operator places the reagent that was just identified. Fail-
ing to do so might lead to wrong results.
If
Do this
There is another bottle
belonging to the set:
A screen is displayed, asking you to scan the bottle.
1. Scan the bottle
2. Insert the bottle.
3. Press
to confirm the insertion.
All bottles of the set are
inserted:
1. Close the main cover.
On the screen with the reagent set buttons, the button
for the new set is now present.
A
Status description
Figure B-7
A
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...