8 General maintenance
Maintenance actions
Clean and disinfect housing
f
Utilities
>
Maintenance
>
Clean Housing
You must clean the housing regularly to ensure that contaminants do not build up
and affect the efficiency of the instrument as a measuring system.
Operator time
Approximately 10 minutes.
Prerequisites
Instrument is switched off.
Tools and materials required
m
Ethyl alcohol
m
Tissues
a
To clean the housing
1
Make sure the system is in
Standby
status.
2
Shut down the system and switch off the instrument.
3
Wipe all the outside of the instrument with alcoholic cleansing tissues.
4
Open the main cover and the right service flap.
5
Wipe all the inside of the cover and flap with alcoholic cleansing tissues.
6
Close the right service flap and then the main cover.
7
Push the transfer head to the rightmost position.
8
Open the main cover and the left service flap.
9
Wipe all the inside of the flap with alcoholic cleansing tissues.
10
Close the left service flap and the main cover.
11
Push the transfer head to its home position.
12
Remove the transfer head cover.
(Press the release buttons on both sides and lift.)
13
Wipe all the inside of the cover with alcoholic cleansing tissues.
14
Reinstall the transfer head cover.
(Push down until the release buttons latch on.)
15
Remove the left and right side panels.
16
Wipe the panels with alcoholic cleansing tissues.
17
Reinstall the left and right side panels.
Make sure that you have read and understood section
The following warning messages in particular are relevant:
o
Injury through working solutions
o
Infection by biohazardous materials
o
To find out what steps are required, you can choose
Utilities
>
Maintenance
>
Clean Housing
and then press
. (The maintenance definition screen is dis-
played.)
o
Do not press
at this stage
. You do this only when you have completed the main-
tenance action. Instead, press
to close the screen.
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...