8 General maintenance
Cedex Bio System
Maintenance actions
Maintenance actions
The following sections describe the maintenance actions of the main instrument.
e
For specific ISE maintenance actions, see
Safety information
WARNING
Injury through working solutions
Direct contact with cleaning solutions or other working solutions may cause personal
injury. When handling such solutions, exercise the precautions required for handling them,
observe the cautions given in the package insert, and observe the information given in the
Safety Data Sheets available for Roche Diagnostics cleaning solutions.
Personal injury or damage to the analyzer due to contact with instrument
mechanism
Do not touch moving parts during instrument operation.
Infection by biohazardous materials
Contact with samples containing material of human origin may result in infection. All mate-
rials and mechanical components associated with samples of human origin are potentially
biohazardous.
o
Be sure to wear protective equipment. Take extra care when working with protective
gloves; these can easily be pierced or cut, which can lead to infection.
o
If any biohazardous material is spilled, wipe it up immediately and apply disinfectant.
o
If waste solution contacts your skin, wash it off immediately with water and apply a dis-
infectant. Consult a physician.
Incorrect results due to build-up of contaminants
During use, contaminants may adhere to the probe. As a result, traces of analytes or
reagents may be carried over to the next. Make sure to perform the probe maintenance
actions as soon as they are due in order to prevent potentially false results.
Incorrect results or damage to the analyzer due to dust and soiling
The user can leave the main cover open while the system is in
Standby
status or while the
instrument is shut down. This can cause dust and dirt being collected in the heating chan-
nel, which in turn might decrease the quality of the cuvettes.
Keep all covers closed. Open them only to perform operation actions.
Incorrect results or processing stop due to skipping maintenance actions
Not performing maintenance actions that are due may lead to situations where the system
cannot continue processing orders, or it may lead to incorrect results. If at all possible, per-
form the maintenance actions when they are due.
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...