Roche Diagnostics
A-90
Operator’s Manual · Version 3.1
4 Software
Cedex Bio System
Key screens
Disk and reagent status
f
Overview
>
The color of a reagent set button represents the status of the set:
The reagent set is on board, but it is blocked for one of the following reasons:
o
The number of available tests is 0.
o
The set is incomplete.
o
The test needs calibrating.
There are fewer than 10% of tests are left for this set.
The expiration date has passed.
The reagent set is on board and ready for use.
o
There is no application that uses this reagent set.
o
A required diluent or cleaner is not on board.
Display detailed information on the status of this test.
Load a reagent set. The number to the left of the button indicates the number of free
positions on the disk.
Display a list with all tests on board, together with information on their status.
On the list, the following abbreviations are used to indicate the status of the reagent
set:
o
C
: Calibration missing
o
E
: Empty
o
I
: Incomplete
o
N
: Not used
o
L
: Low
o
X
: Expired
A
A plus (+) indicates that there is already an active identical reagent set on board. (This icon is
displayed as soon as the first bottle of the set is loaded.)
B
The number to the left of the button indicates the number of free positions on the disk.
Figure A-56
Reagent sets loaded on the reagent disk
A
B
<xyz>
5
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...