Roche Diagnostics
Operator’s Manual · Version 3.1
E-11
Cedex Bio System
12 ISE description
Hardware
ISE tower
A hollow, transparent cylinder with multiple inlets and outlets for air and solutions.
In the ISE tower, samples and diluents are mixed using an air stream.
Tubing
Tubing is used for transporting the fluids. It connects the ISE unit with the water and
waste system, and also with the ISE Reference Solution and ISE Calibrator
indirect/urine bottles, which are placed on the ISE unit.
Sensors
The ISE sample sensor detects the presence of liquid (sample, control, standard
solution, or ISE Calibrator indirect/urine) or air and enables the correct placing of
segments for measurement and cleaning.
The ISE Reference Solution sensor detects the presence of ISE Reference Solution.
ISE fluid bottles
Two bottles are placed on the ISE unit:
o
ISE Calibrator indirect/urine bottle (blue label)
o
ISE Reference Solution bottle (red label)
Level monitoring is performed on the basis of weight, the bottles are placed on scales.
Fluid containers
The following table shows which container is used for which fluid:
Fluid
Container(s)
Position
Samples
Tube
Sample area
Controls
Tube
Sample area
ISE Solution 1 and 2
Tube
Sample area
Cleaner
External cleaner bottle
External fluid rack
ISE Deproteinizer
Tube
Sample area
Activator
Tube
Sample area
ISE Etcher
Tube
Sample area
ISE Reference Solution
Bottle
ISE unit
ISE Calibrator indirect/urine
Bottle
ISE unit
Water
External water container
External fluid rack
Waste
External waste container
External fluid rack
Table E-2
Fluids, containers, and where they are used
The term tube includes all kinds of tubes, as long as their dimensions lay within prescribed
limits. It also includes secondary tubes (cups). See
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...