9 Messages and alarms
Cedex Bio System
List of alarm messages
5044
[%s] Error main board fan. The fan could not be started or
the fan speed is not correct.
Restart the system. If the error persists call service.
5047
[%s] Error cooling fan. The fan could not be started.
Restart the system. If the error persists call service.
5049
[%s] Error sample area fan. Switch off the instrument.
Remove the right side panel and check if the fan is jam-
ming or blocked. If possible eliminate the problem.
Restart the system. If the error persists call service.
5052
[%s] Error cooling system. The cooling heat sink
overheated.
Ensure unrestricted cooling air circulation. Check the
cooling fan and make sure the external vent is not
blocked. Restart the system. If the error persists call ser-
vice.
5054
[%s] The analyzer heating system overheated.
Switch off the instrument and wait 15 minutes to let cool
down the analyzer. Restart the system. If the error persists
call service.
5055
[%s] The analyzer temperature is outside the range.
Restart the system. If the error persists call service.
5056
[%s] The calibration of the transfer head fluid sensor
failed.
Ensure that the sample probe tube is correctly installed.
Ensure that the syringe is correctly installed. Restart the
system. If the error persists call service.
5058
[%s] Error fluid sensor transfer head. The sensor does not
react anymore.
Restart the system. If the problem persists, call service.
5060
[%s] A not specified hardware error (%d) was reported by
the Multi-Slave control.
Restart the system. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
5061
[%d] Air detected in the fluid system.
Refill the water container. Check the filling level of the
cleaner bottle. Replace the cleaner bottle if required. Per-
form <Prime Fluid System>. If the error persists call ser-
vice.
5062
The degasser fluid sensor is not initialized.
Perform <Initialize Degasser Fluid Sensor>. Follow the
instructions provided on the display. If the error persists
call service.
5063
[%s] Air bleed valve error. The air bleed valve does not
react anymore.
Restart the system. If the error persists call service.
5064
[%s] Error fluid sensor degasser. The sensor does not
react anymore.
Restart the system. If the error persists call service.
6001
[%s] Photometer error. Signal is invalid or unstable.
Switch off the instrument. Remove and clean the rotor. In
case of a recurrence perform <Replace Photometer
Lamp> and <Abs Air/Water Calibration>. If the error
persists call service.
6002
[%s] Photometer error. The photometer has a wrong
status because initialization has not been successfully
performed yet.
Restart the system. If the error persists call service.
6018
[%s] A not specified photometer hardware error (%d) was
reported by the photometer.
Restart the system. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
6019
[%s] Photometer error. The photometer signals were
outside the range while performing a photometer
initialization or a photometer measurement.
Restart the system. If the error persists call service.
ID
Message
Comment
Table D-1
List of alarm messages (Continued)
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...