ISE maintenance actions
To remove a pinch valve cap and clamp:
o
Push the cap horizontally in the direction of the arrow, away from the cut-out.
o
Lift and put it aside.
o
Lift the valve clamp and put it aside.
Disconnecting the tubing from
ISE tower
11
Disconnect the tubing from the ISE tower
o
Remove the ISE overflow collector (B) by pulling it upwards and turning it
from side to side as you do so.
Place the collector behind the wash station.
o
Remove the locking ring (C) from the base of the ISE tower by turning it
counter-clockwise one quarter of a turn and then lifting it over the ISE tower.
o
Lift the ISE tower (A). Check whether any O-ring is stuck to the ISE tower. If
so, remove it and place it back in its position on the ISE tower base.
o
Disconnect the tubing.
B
C
A
Do not remove the piece of tubing at the underside of the mixing tower. It will be
changed when a service representative performs periodic maintenance.
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...