Roche Diagnostics
B-72
Operator’s Manual · Version 3.1
5 Daily operation
Cedex Bio System
Performing QC
5
Press a control button.
A screen is displayed, asking you to place the selected control.
6
Place the control tube on any free position on the sample area.
The system registers the position and automatically defines an order.
If there are more controls to place, the screen with the control buttons is displayed
again. The control you just placed is no longer active.
7
Select and place the remaining controls as described in steps 5 through 6.
When all controls have been selected and placed, the
Overview
tab is displayed.
8
Press
to start processing the control orders.
Validating QC results
Flagged QC results
If flags are generated, you must determine their cause and decide whether to accept
the result or to ignore it.
The effect of flagged QC results depends on the configuration.
e
Editing the acceptable flags list
Configuration setting
Effect
QC Auto Accept
:
On
Results that do not contain a flag are auto-
matically accepted. Results with flags that are
marked in a predefined list of flags that
should be ignored are also automatically
accepted.
QC Auto Accept
:
Off
All results need to be manually accepted.
QC Acceptable Flags
Flags that are marked in this list will be
ignored by the system.
Table B-8
QC flag configuration and its effect
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...