Roche Diagnostics
B-160
Operator’s Manual · Version 3.1
7 Configuration
Cedex Bio System
Configuration
Editing the acceptable flags list
When working with automatic accepting of results, results that are not flagged are
automatically accepted. Results with flags that are marked in an acceptable flags list
are accepted as well. There is such a list for sample, QC, and calibration results.
a
To add or remove a flag from an acceptable flags list
1
Choose
Utilities
>
Configuration
>
Result Handling
.
2
Expand the list.
3
Select one of the following entries:
o
Sample Accept. Flags
o
QC Acceptable Flags
o
Cal Acceptable Flags
4
Press
.
A list of relevant flags is displayed.
5
Press the square to the left of a flag name to select it or to cancel its selection.
A selected flag shows a check mark (
l
).
6
Press
.
Mapping the host codes
Applications are delivered with a unique ID called code. Laboratories often use their
own codes for applications. By mapping the codes used by the laboratory (host codes)
to those defined by the application manufacturer, the laboratory can work with their
own accustomed codes. The
Cedex Bio
instrument provides a mapping table that can
easily be edited.
Editing a mapping table involves the following steps:
1. Exporting the mapping table to an USB stick. (The table is written as a text file.)
2. Editing the file on a computer and save it as a text file to the stick.
3. Importing the file to the
Cedex Bio
instrument.
a
To export the mapping table
1
Choose
Utilities
>
Applications
>
Host Codes
.
2
Press .
A screen is displayed, asking you to insert the USB stick.
3
Insert the USB stick.
4
Press .
5
Select the directory where you want to save the mapping file.
6
Press
to confirm the selection.
The file is exported as a text file of the name format hct_yyyymmddhhmmss.txt.
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...