Roche Diagnostics
B-76
Operator’s Manual · Version 3.1
5 Daily operation
Cedex Bio System
Performing QC
3
Press
to display the results in a table.
The results are sorted chronologically. An overview is followed by statistical data
for all stored results and for the results of the current month. (If a QC target value
change took place, only the results from this date onward are taken into account
for both statistics.)
4
Press
to close the screen.
Performing reagent mixing
Reagents that contain Latex granules, for example D-Dimer, require periodic mixing.
The mixing interval is part of the application definitions and cannot be changed by
the user. The system checks every 30 minutes for reagent sets that require mixing.
e
For mixing during the
Prepare
phase, see
a
To perform the mixing
1
Choose
Overview
>
.
2
Press the button of the reagent set you want to mix.
A screen is displayed that contains details on the set.
3
Press .
(This button is available if there is a mixing interval defined for this reagent set.)
Mixing starts. A screen is displayed informing you about the progress of the
mixing action.
4
When mixing is finished, press
to close the screen.
Closing the dialog box while mixing is in progress stops the mixing action.
Mixing would have to be performed anew.
A
Date of QC measurement
B
Result or event
C
Flag (if generated)
Figure B-29
A
B
C
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...