B-29
Cedex Bio System
5 Daily operation
Preparing the system
Replacing cuvette segments
2
Press a segment button.
A screen is displayed, informing you that the system is ready for cuvette handling.
3
Wait for the main cover LED to turn green.
4
Open the main cover.
5
Remove the segment and treat it as biohazardous waste.
6
Insert a new cuvette segment.
7
Press
to confirm the replacement.
(Press
if you removed the segment without replacing it.)
8
Do one of the following:
9
When you have finished replacing cuvettes, press
to proceed to the next
stage in the
Prepare
wizard.
Performing reagent mixing
Reagents that contain Latex granules require periodic mixing. The mixing interval is
part of the application definitions and cannot be changed or deleted by the user. The
system checks every 30 minutes for reagent sets that require mixing.
Within the
Prepare
phase a screen is displayed that lists all reagent sets for which a
mixing interval is defined.
a
To perform the mixing
1
Select the reagent sets you want to mix.
If working with the
Prepare
wizard, the reagent sets that need mixing are
automatically selected.
2
Press .
Mixing starts. A screen is displayed informing you about the progress of the
mixing action.
(If there is more than one bottle in the set with a mixing interval defined, it will be
mixed as well, irrespective of being due or not.)
3
When mixing is finished, press
to close the screen.
WARNING
Incorrect results due to scratched or soiled cuvettes
Scratches and impurities on the cuvettes distort the measurements.
Do not touch the cuvettes and make sure they do not touch other items when handling
them.
If
Do this
You want to exchange another segment:
This was the last segment you wanted to
exchange:
Close the main cover.
When handling cuvette segments without using the Prepare wizard:
Choose
Overview
>
. The system must be in
Standby
status.
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...