Roche Diagnostics
C-20
Operator’s Manual · Version 3.1
8 General maintenance
Cedex Bio System
Maintenance actions
6
Replace the water inlet filter.
o
Remove the retaining ring by pulling it by its tab away from the filter casing
(A, D, E).
o
Remove the filter (B) and replace it with a new one.
o
Place the retaining ring.
Press it firmly on the filter casing by placing your thumb across it an applying
equal pressure all around.
7
Reinsert the tubing adapter into the water container. Press down firmly.
8
Press .
The maintenance action
Prime Fluid System
is automatically performed.
A message will inform you when the maintenance action is complete.
9
Press .
A
Removing the retaining ring
B
Filter
C
Filter casing
D
Retaining ring
E
Tab on retaining ring
Figure C-2
A
C
B
E
D
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...