Roche Diagnostics
B-130
Operator’s Manual · Version 3.1
6 Special operations
Cedex Bio System
Connecting and disconnecting the external fluid containers
a
To disconnect a tube from the instrument
1
Remove the right side panel.
2
Do one of the following:
If
Do this
You want to disconnect the water tubing:
Press down the release clamp (A) on the
connecting socket and pull the connector
away from the socket.
You want to disconnect the cleaner tubing:
1. Pull out the protective sleeve (B) and
move it along the tube far enough that
you can grip the connecting screw.
2. Turn the connecting screw (D)
counterclockwise until the connector is
released.
You want to disconnect the waste tubing:
Press down the release clamp (C) on the
connector and pull the connector away
from the socket.
A
Release clamp on water connecting socket
B
Cleaner connector with protective sleeve
C
Waste connector release clamp
D
Cleaner connector screw
Figure B-46
A
B
C
D
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...