6 Special operations
Adjusting the touchscreen
Adjusting the touchscreen
With a touchscreen, it is important that the point where you press the screen
corresponds exactly with its hardware equivalent. If this were not the case, pressing a
screen item such as a button might not lead to the expected result.
a
To adjust the touchscreen
1
Press the
global button.
A screen is displayed that offers several kinds of stopping. Ignore it.
2
Press the
global button again.
A screen for adjusting the screen is displayed.
3
Press exactly in the center of the black cross.
When you have touched it properly, it turns yellow and the next cross turns black.
4
Do the same for all crosses.
5
When all crosses are yellow, a message is displayed, informing you that you need
to reboot the system before the new calibration values take effect.
6
Press
to close the calibration screen.
7
Restart the system.
o
Press
.
o
Press
Restart
.
The system shuts down and automatically restarts.
o
Wait until the
Overview
tab is displayed and the system is in
Standby
status.
Cleaning the touchscreen
Because the touchscreen gets easily soiled you should clean it regularly.
Tools and materials required
m
70% ethyl alcohol
m
Tissues
m
Protective gloves
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...