E9
An accurate determination of the scale of the figure
can be performed with an object micrometer:
(a) make a photomicrograph
(b) measure the image on the film
Remote control
The Jack in the rear of the control unit is intended for
the connection of a remote control (e.g. foot button),
which takes over the function of the START button.
Micro Flash Terminal
This Jack is for connecting the connection cable to
the microflash power supply. The shutter is
synchronized as an x-contact for all exposures up to
the shortest of 1/125 sec. The operation is described
in a separate manual.
Brightness-dependent voltage
According to the brightness of the microscopic
image, the output voltage range is 0-5V (for high to
low light intensity).
Magnification in the film plane
Photomicrograph image scale (M) is calculated
according to the formula:
M = Vobj * q * Vok * P
Vobj................................Magnification of the lens
q.....................................Relay magnification
Vok......................Enlargement by the Photo eyepiece
P......................................Camera factor
The table below provides information on the product
Vok * p, considering the camera format and the
positions L or H of the magnification changer in the
photosystem
Camera
Format
Vok * P
L H
24 x 36
1.8
2.9
3 1/4” x 4 1/4”
5
8
4” x 5”
9 x 12cm
6.3
10
Summary of Contents for Univar
Page 1: ...Reichert Univar Manual...
Page 2: ......
Page 48: ...D6 blank no content...
Page 58: ...E2 blank no content...