Lamp House 25
The lamp house 25 is manufactured in two versions; one
mounted on the left side of the mirror house, the other on
the right side. The illustration shows the left-mounted
version.
The following lamp can be used in the lamp house 25:
HBO 200W/4 or CS 200W/4 mercury vapor high-pressure
lamp; CSI 250 W/1 metal halide short-arc lamps
Use of mercury-vapor, high-pressure lamp HBO
200W/4 or CS 200W/4 and the CSI metal halide
short-arc lamp 250W/1
PLEASE NOTE!
Disconnect the connection cable
➀
before opening the
lamp house!
⚫ Handle the lamp carefully to prevent mechanical
damage. Touch the lamp base instead of the quartz body.
When tightening the lamp, always hold the base so no
rotation or bending forces act on the quartz body. Before
first use, clean the quartz body with a clean cloth and
alcohol, since fingerprints on the quartz body will burn.
For reasons of operational safety, change the lamp when
the average lifespan is reached – please observe the
instruction manual for the lamp.
⚫ With the coin screw
➁
, open lamp house. Screw the
lamp into the bracket
➃
, so that the inscription on the
base is legible. If the arc point of the quartz bulb is not be
located in the optical beam path (reflector-collector axis)
when the lamp housing is closed, loosen the grub screw
➄
to correct the position of the lamp. Then tighten the
grub screw
➄
.
⚫ Connect the cable
➂
to the threaded stud of the upper
lamp end and tighten with the knurled nut.
⚫ Close the lamp housing and tighten the coin screw
➁
.
⚫ Electrical connections for the HBO 200W/4 or CS
200W/4 mercury-vapor high-pressure lamps and the CSI
250 W/1 metal halide short arc lamp: the electrical
connection of the lamp is via the connecting devices
described on page D12.
PLEASE NOTE!
Close the lamp house before lighting the lamp, to protect
the sensitive electronics in the photo automation circuitry