D12
Connections for HBO 200W/4 or CS 200W/4
mercury-vapor high-pressure lamp
1. safety starter
2. hour counter
3. power switch
4. connector for lamp 25
The power supply is intended for 220VAC. If the line
voltage is not 220V, the intermediate transformer,
page D13, is connected upstream. The housing has
a socket for the line power connection on the back.
⚫ By default, the power supply is set for a line
frequency of 50Hz and an L1 lamp. There is a
selector option for L2 lamps at 50Hz or L1 lamp at
60 Hz.
⚫ When turning the lamp on, please refer to page
D7!
⚫ The lamp is turned on with the power switch
➂
.
When the ignition is completed, the glow lamp of the
safety starter
➀
goes out. If the lamp does not
ignite, the red button of the safety starter pops out
to protect of the ignition device. The red button
must be pushed in again (reset) before again turning
on with the switch
➂
. It is also advisable to take a
break before re-igniting, so that the lamp can cool
down.
⚫ The burning time can be read on the hour counter
➁
. Please pay close attention to the lamp
manufacturer specifications.
Power supply for CSI 250W/1 metal halide
short-arc lamps
5. switch
6. Starter
7. connector for lamp 25
⚫ The power supply is suitable only for alternating
current. It is set for a 220V/50Hz standard and can
be switched to 220V/60Hz or 240V/50Hz. If the line
voltage has a different value, the intermediate
transformer, page D13, should be used upstream.
The housing has a socket for the line power
connection on the back.
⚫ Please follow the notes on page D7 when the
lamp is switched on!
⚫ To turn on the lamp, first turn on the switch
➄
and then push and hold the starter
➅
for about 2-3
seconds. The light in the starter must light up while
the lamp is lit.