E4
⚫ The connection cable of the automatic camera
with the multiple plug is connected to the associated
connector at the rear of the microscope. There are
three female connectors: one for a camera on the
left, the second for a camera on the right, and the
third for a top-mounted automatic camera.
⚫ Please make sure that the unused sockets are
provided with shorting plugs!
⚫ Operating voltage: the control unit is normally set
to 220V. When switching to a different voltage, the
inner head of the voltage selector is pushed in and
removed after a quarter-turn of the screw
counterclockwise. With a coin or screwdriver blade,
you can then move the voltage selector to the
desired voltage.
⚫ For voltages up to 150V, one changes the internal
fuse (0.25A) for one rated 0.5A.
⚫ The control unit is then connected to the power
supply with the supplied cable.
Please note!
For thermal reasons, the control unit for the camera
system should always be located at the bottom,
when several control-boxes/power-supplies are
placed one above the other.