Instruction LSM880 short 2017-12-26.docx    2017-12-26

 

 

 

LSM 880 quick operation manual 

Switch on

 

-

 

Make sure room temperature is 22-24°C. 

-

 

If necessary turn on air condition 

-

 

In the main switch box turn on the following switches 

 2 

 3 (do not touch the key) 

-

 

Turn on the computer and log in 
User name: multilabs 
Password: 123456 

Position 

Objective lens

 

Magni-

fication 

NA

 

coverslip

 

Working 
distance 

Immersion 

Parfocal length 

Plan Apochromat

  

x10

 

0.45

 

0.17 mm 

2 mm 

Air 

45 mm 

Plan Apochromat

 

x20

 

0.8

 

0.17 mm 

0.55 mm 

Air 

45 mm 

4 (bombila) 

W Plan Apochromat

 

x20

 

1.0

 

1.8 mm 

Water dipping 

75 mm 

1+ adaptor 

W Plan Apochromat

 

x40

 

1.0

 

2.5 mm 

Water dipping 

45 mm 

1+ adaptor 

Plan Apochromat

 

x63

 

1.4

 

0.17 

0.19 mm 

Oil

 

45 mm 

Software and acquisition  

-

 

Log into BookitLab and activate your reservation to start Black Zen 2.3

 

-

 

To acquire images click Start System

 

-

 

In the acquisition tab 

 Laser tab 

Switch on the lasers, (The Argon laser should be turned on by Standby 

 ON. Please wait 5' minutes until 

the laser stabilizes, its status should be optimal in the properties tab). 

-

 

Sample mounting and viewing 
Insert the slide or dish in the holder. 
To change objective turn the black wheel to point up to release the objective turret. Turn manually to the 
required objective, lower the objective

 

by pointing the black wheel in the objective turret down, (if an 

objective needs to be replaced please ask the microscopy staff). 
In the software go to Locate tab. 
On the microscope’s right side choose VIS  

-

 

Transmitted light 
Click the transmitted light lamp icon, click ON and set desired brightness (or via the microscope). 
Set the filter turret position to DIC and the condenser below the stage to II or III according to the objective 
lens (see objective lens info on touchpad).

 

Set up Kohler illumination.

 

-

 

Reflected light (fluorescence) 
Open fluorescence shutter

 

Set the filter turret to the desired position 

-

 

Image Acquisition

 

On the microscope’s right side choose LSM

 

In Zen, Acquisition tab: 

 

Load a saved configuration from the Experiment Manager or open an image and below the image click 
Reuse

 

Select Show all tools and open the following tabs:

 

Reviews: