Light Path, Channels, Acquisition Mode

 

With one track checked, in fast settings (512x512, speed 8, no averaging) click Continuous to focus. 
Optimize each channel separately by checking Range Indicator and setting pinhole (1 Airy unit) detector 
gain, laser power, offset, digital gain.

 

Check all channels and set acquisition parameters: zoom, pixel number, scanning speed, averaging. 
After setting zoom press “optimum” so that the system will calculate number of pixels according to the 
optimal Nyquist number for the given objective. If scanning time is too high you can consider saving time 
by:  

 

lowering pixel dwell time.  

 

setting scanning direction to meander instead of comb 

 

decreasing number of averaging iterations 

 

using frame instead of line mode

 

Press Snap to acquire image 

 

For multidimensional images check the desired dimension (Z, time, tile) and after setting the relevant 
parameters click Start Experiment

 

See instructions booklet for more details 

-

 

Saving images 

 

Save image as CZI filetype in D:\LSM 880 USERS DATA\PI name\user name\mm-yyyy

 

 

-

 

Data backup

 

Data are copied automatically every night, if computer is on and connected to the internet.

 

PLEASE do not use any type of USB flash drive to retrieve your data.  

There are computers with active scanning anti-virus software at the Computerized Analysis Room, 
connected to the servers.

 

 

Ending session 

 

Before you finish please make sure the next user is coming. 

 

If yes:

 

Close the ZEN software

 

Log off your reservation at BookItLab

 

clean oil immersion objectives with lens tissue dipped in petroleum ether, twice 

clean water dipping objectives by dipping in DDW (e.g., in 35mm dish) 

Leave nosepiece at 10x objective lens

 

Center the stage

 

Leave surfaces clean 
 

If no user is coming in the next two hours, also, in this order:

 

Turn off the lasers in the software 

Close the ZEN software

 

Log off your reservation in BookItLab

 

Switch off 

 2 in the switch box

 

Wait until the Lasos fan shuts off and only then switch off 

 1 

Cover microscope 

Turn off air condition

 

Do not turn off computer unless you are the last user before the weekend.  

Please do not use any type of USB drive to copy your data.

 