![background image](http://html1.mh-extra.com/html/rands/nrx-b4/nrx-b4_user-manual_3240273045.webp)
Measurement Basics
R&S
®
NRX
45
User Manual 1178.5566.02 ─ 05
Example: The sensor does not support the measurement type
The notification center indicates an error, but the no control element is highlighted.
The error message reports a settings conflict.
Solving the error
1. Tap the
displayed measurement value or graphic
.
The "Measurement Settings" dialog is displayed.
The "Measurement Type" is highlighted.
2. Select another measurement type that the sensor supports, or assign another sen-
sor.
A setting that differs from the preset value is also indicated across the hierarchies by a
pencil symbol, if the visualization is enabled. See
Settings Conflict