Remote Control Commands
R&S
®
NRX
305
User Manual 1178.5566.02 ─ 05
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
Range:
0.0 to 15.0
*RST:
0.0
Default unit: s
Manual operation:
See
"Exclude from Start, Exclude from End"
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:DCYCle:STATe
<state>
Effective for continuous average measurements.
Enables or disables the duty cycle correction.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<state>
*RST:
OFF
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:DCYCle[:VALue]
<duty_cycle>
Effective for continuous average measurements.
Using the duty cycle correction, the average power of RF bursts is calculated from the
average power of the whole signal. Essentially, the average power of the whole signal
is divided by the set duty cycle.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<duty_cycle>
Range:
0.001 to 100.0
*RST:
50.0
Default unit: pct
Manual operation:
See
Configuring Sensors