Remote Control Commands
R&S
®
NRX
342
User Manual 1178.5566.02 ─ 05
OUTPut:RECorder<output>:LIMit:UPPer:RATio:RLOSs
<value>
Enter the return loss measurement value that corresponds to the output voltage of
2.5
V. Above this limit, the output is set to 2.5
V.
Suffix:
<output>
.
1 to 2
BNC connectors at the rear; 1
=
Out
1
/
Trig
Out,
2
=
Trig
In
/
Out
2
Parameters:
<value>
Range:
-180.0 to 180.0
*RST:
10.0
Default unit: dB
Manual operation:
See
OUTPut:RECorder<output>:LIMit:UPPer:RATio:SWR
<value>
Enter the SWR measurement value that corresponds to the output voltage of 2.5
V.
Above this limit, the output is set to 2.5
V.
Suffix:
<output>
.
1 to 2
BNC connectors at the rear; 1
=
Out
1
/
Trig
Out,
2
=
Trig
In
/
Out
2
Parameters:
<value>
Range:
-1e18 to 1e18
*RST:
10.0
Default unit: -
Manual operation:
See
OUTPut:RECorder<output>:LIMit:UPPer:RATio[:VALue]
<value>
Enter the power ratio measurement value that corresponds to the output voltage of
2.5
V. Above this limit, the output is set to 2.5
V.
If you enter a value without unit, the unit is defined by
. For further information, see
Suffix:
<output>
.
1 to 2
BNC connectors at the rear; 1
=
Out
1
/
Trig
Out,
2
=
Trig
In
/
Out
2
Parameters:
<value>
Range:
-180.0 to +180.0
*RST:
+10.0
Default unit: dB
Manual operation:
See
Configuring the Analog Signal Ouput and the Trigger Output