Remote Control Commands
R&S
®
NRX
311
User Manual 1178.5566.02 ─ 05
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:OFFSet:TABLe[:
STATe]
<state>
Enables or disables the frequency-dependent offset correction specified in the selected
table.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Setting parameters:
<state>
*RST:
OFF
Usage:
Setting only
Manual operation:
See
"Frequency dependent offset active"
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:OFFSet[:
MAGNitude]
<value>
Sets an offset correction value for the selected reference plane. This can correct the
influence of cables etc. and does not influence measurement accuracy.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<value>
Range:
-200.0 to 200.0
*RST:
0.0
Default unit: dB
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:SGAMma:CORRection:STATe
<value>
Enables or disables gamma correction in order to achieve higher measurement accu-
racy.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Configuring Sensors