Remote Control Commands
R&S
®
NRX
188
User Manual 1178.5566.02 ─ 05
DISPlay[:WINDow<Window>]:ANALog:UPPer
<value>
DISPlay[:WINDow<Window>]:METer:UPPer
<value>
Sets the upper limit value of the analog scale.
Suffix:
<Window>
.
1 to 4
Measurement channel
Parameters:
<value>
See
DISPlay[:WINDow<Window>]:METer:LOWer
[SENSe<Sensor>:]TRACe:OFFSet:TIME
<time>
Sets the duration of the trace.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<time>
Range:
-15.0 to 15.0
*RST:
0.0
Default unit: s
Manual operation:
See
[SENSe<Sensor>:]TRACe:TIME
<time>
Sets the vertical scaling.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<time>
Range:
8.3e-9 to 30.0
*RST:
0.01
Default unit: s
Manual operation:
See
14.4.1.3
Units
If you enter a value that is expressed in a certain unit, for example Hz, you can omit
the unit. Then, the default unit provided in the remote command description is used. If
you need decimal multiples and submultiples of a unit, you can use SCPI prefixes, see
Measurement Settings and Results