Remote Control Commands
R&S
®
NRX
325
User Manual 1178.5566.02 ─ 05
[SENSe<Sensor>:]BANDwidth[:RESolution]:TYPE:AUTO[:STATe]
<state>
Only effective for R&S
frequency selective power sensors.
If enabled, sets the filter type suitable for the currently chosen measurement mode and
bandwidth.
You can query the selected filter type using
.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<state>
*RST:
ON
[SENSe<Sensor>:]BANDwidth[:RESolution]:TYPE[:VALue]
<value>
Only effective for R&S
frequency selective power sensors.
Sets the filter type for resolution bandwidth filter. The filter bandwidth is not affected.
If you want to set the filter type automatically, use
.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<value>
FLAT | NORMal | LTE | W3GPp
*RST:
FLAT
[SENSe<Sensor>:]BANDwidth[:RESolution][:VALue]
<value>
Only effective for R&S
frequency selective power sensors.
Sets the resolution bandwidth.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<value>
Range:
10.0 to 400.0e6
*RST:
25.0e6
Default unit: Hz
Configuring Sensors