Remote Control Commands
R&S
®
NRX
332
User Manual 1178.5566.02 ─ 05
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO[:
STATe]
<state>
If enabled, determines the average count automatically from the level of the input sig-
nal.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<state>
*RST:
ON
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt[:VALue]
<value>
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:
OFF
is set.
Sets the number of readings that are averaged for one measured value. The higher the
count, the lower the noise, and the longer it takes to obtain a measured value.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
Range:
1 to 1048576
*RST:
4
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:APERture:MODE
<mode>
Selects a user-defined or default value for the integration time.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<mode>
DEFault | USER
*RST:
DEFault